How to use pH meter?

As you may know, there are few methods to measure pH. These include two traditional methods such as pH paper and pH-sensitive indicator. However, to measure pH for analysis, pH need to be determined quickly and accurately.

This is why electrochemical measurement or pH meter is the most preferred method. It is accurate and easy to operate. The meter measures the slight voltage differences between a reference electrode and a measuring electrode. This small voltage (mV) is then converted to a pH reading by referring to standard slope (pH vs mV).

    There are different designs of pH meter available on the market such as portable, benchtop and integrated sensors. Nevertheless, the operational method stays more or less the same.

The rule of using pH meter is really straight-out simple
  1. Calibration point must be within your sample range. For example, by any reference your sample is supposed to be around pH 6.5 to pH 8.9, choosing the calibration points of pH 4.01, 7.00 and 10.01 are well enough. Calibration using these three points will allow you to read a sample range from about pH 4.01 to pH 10.01 accurately. If your sample is more acidic down to pH 2 then you surely need to include calibration points pH 1.68 and pH 12.45 as well.

  2. Automatic temperature compensation (ATC) probe (if applicable) must be submerged in the buffer or sample during calibration and real measurement of sample. This is very crucial to avoid temperature error associated with pH electrode. Without the use of ATC probe, the measurement will not be accurate when sample deviate from pH 7 and 25°C. 

  3. The electrode must fit with the nature of the sample. Sample with high total solids such as wine and juice required special electrode that is equipped with clogging prevention system. Normal pH probe will be easily contaminated, damaged or inaccurate when used for such sample.
According to theoretical ideal slope, the meter should read 0 mV at pH 7. The mV vs pH graph shows a straight line with positive slope. However, if the line shifted downward or upward, it is called offset. If the offset is too big then you may need to change your electrode. More on that in next article. 